Evaluation of a Multi-Mode ODS Column for the Separation of Polar Metabolites


The Application Notebook


Figure 1
The analysis of polar, ionic metabolites is important for drug discovery and biomarker research. Often times, it is necessary to separate these metabolites from each other and biological matrixes prior to MS detection (1).

Experimental and Results


Figure 2
All data was generated with semi-micro HPLC system equipped with UV or ELS detection. Solutes were separated using LC–MS compatible conditions on Unison UK-C18 (conventional ODS) and Scherzo SM-C18 (Multi-Mode ODS consisting of ODS + anion + cation ligands, Figure 1). Figure 2 shows the separation of a steroid hormone from its metabolites. 17b-estradiol (neutral) showed similar retention on both phases — indicating reversed-phase is the main mode of retention. The glucuronide and disulfate metabolites are poorly retained on conventional ODS, but show improved retention on Scherzo SM-C18 (due to anion exchange). Figure 3 shows the separation of mevalonic acid and mevalonolactone. Mevalonic acid shows improved retention on Scherzo SM-C18 due to anion exchange. Figure 4 shows the separation of glutathiones (both reduced and oxidized forms). These compounds are difficult to separate on conventional ODS, but show improved retention/separation on the Multi-Mode ODS (cation + anion exchange).


Figure 3

Conclusion


Figure 4
Scherzo SM-C18 (Multi-Mode ODS) shows improved selectivity for polar metabolites over conventional ODS. This column should be useful for scientists who work with polar metabolites, as well as those who require an ODS column with different selectivity.

References




(1) "Using Mass Spectrometry for Drug Metabolism Studies;" Korfmacher, W.A., Ed.; CRC Press: Boca Raton; Fl, 2005.

Imtakt USA (formerly Silvertone Sciences)
1511 Walnut St., Suite 310, Philadelphia, PA 19102
tel. (215)665-8902; fax (501)646-3497
Email:
;
Website: http://www.imtaktusa.com/

ADVERTISEMENT

post a comment
Your email address will NOT be published.
appears with your comment
read our privacy policy
Note: does not support HTML
All Comments submitted are subject to review, and may be delayed before posting. Comments will be posted at the discretion of the editor.
LCGC E-mail Newsletters
Select each newsletter you wish to subscribe for:
| Monthly
| Weekly
| Monthly
 Name:
 E-mail:
more info
Survey
Which stationary phase do you use in RP chromatography?
C18
C8
C4
C1/C2
Phenyl
Cyano
Other
C18
80%
C8
9%
C4
1%
C1/C2
1%
Phenyl
6%
Cyano
3%
Other
0%
View Results
Source: The Application Notebook,
Click here