 John W. Dolan
|
Here we are, more than 40 years after the initial exploration of high performance liquid chromatography (HPLC). Many things
have changed, some innovations have come and gone, but the reciprocating-piston pump still remains as a key component of the
LC system. We've tried pneumatically driven pumps, piston-diaphragm pumps, syringe pumps, and some other ideas, but none have
proven to be superior to the reciprocating-piston pump. Yes, some changes have been made, but the basic operation remains
the same. This month's "LC Troubleshooting" installment will take a look at the design of these durable pumps and also examine
some of the potential weaknesses and how to overcome them.
The Basic Design
 Figure 1: Single-piston reciprocating pump: (a) Pump in intake, or fill cycle; (b) pump in delivery cycle; and (c) partial
section of piston showing cut-away portion of piston seal. See text for discussion.
|
The basic design and operation of the reciprocating-piston pump is illustrated by the single-piston pump shown in Figure 1.
The key components are a piston, pump seal, pump head, and a couple of check valves. The piston, usually made of sapphire,
is driven back and forth in the pump head by a rotating motor. Various means have been conceived to convert the rotary motion
of the motor into the bidirectional movement of the piston. Most commonly, this is done by a cam pressing against one end
of the piston to push it into the pump head and a spring to push the piston back out. The pump seal is shown in the inset
in Figure 1c. It is a polymer ring that fits around the piston, and a small lip forms a liquid-tight seal against the piston
with the aid of a spring and liquid pressure. A pair of ruby check valves with sapphire seats are mounted on the top and bottom
of the pump head. The check valves control the direction of flow through the pump. On the intake stroke (Figure 1a), the piston
is withdrawn, which creates a low pressure area inside the pump head. This allows the outlet check valve to close and the
inlet check valve to open, so that mobile phase flows in to fill the pump head. On the delivery stroke (Figure 1b), the piston
moves into the pump head, and the inlet check valve is closed as the pressure increases. When the pressure inside the pump
head exceeds the pressure in the column, the outlet check valve opens and mobile phase flows to the column. When all is working
well, this simple pump design is quite reliable.
The weak points in the design are the check valves and the pump seal. The operation of the check valves and some design improvements
were discussed in the June 2008 installment of "LC Troubleshooting" (1). Contamination of the check valves can cause them
to leak, and under the right circumstances, acetonitrile in the mobile phase can cause the inlet check valve to stick closed.
The pump seal forms a seal against the moving piston, and although the sapphire piston is very smooth, the seal eventually
will wear out. This wear is accelerated if buffer is allowed to sit in an unused pump, because the liquid behind the seal
evaporates and leaves an abrasive layer of buffer crystals. These abrade the seal when the pump is restarted, and can shorten
the seal life. Generally the pump seals will last 6–12 months under normal operation if buffers are rinsed from the system
before shutdown. Bubbles of air trapped in the pump head will cause the pump to under-deliver mobile phase. Bubbles can result when mobile
phases are not degassed sufficiently. Fortunately, most of today's LC systems incorporate an inline vacuum degasser to degas
the mobile phase automatically. So if the pump is purged of bubbles when it is started, bubble problems in the pump are not
a common problem today — they were the bane of the chromatographer in years past.
The original reciprocating-piston pumps were crude by today's standards. I remember using the Milton–Roy Minipump, one of
the standard pumps in the late 1960s and early 1970s, particularly for laboratory-built systems that were common in graduate
school laboratories. The flow rate was controlled by adjusting a stop that limited the distance the piston could move. With
a single-piston design, the pump spent half the time filling and half the time delivering mobile phase. This resulted in pulses
of pressure and flow that were reflected in a fluctuating baseline and short column lifetimes of the then-poorly-packed columns.
Huge pulse dampeners, sometimes 50–100 mL in volume, were required to reduce the pulses to an acceptable level. Refinements
in the shape of the driving cam and the use of stepper-driven motors helped to minimize pulses with single-piston pumps, but
as other aspects of the LC system were improved, single-piston pumps were not sufficiently pulse-free to be acceptable.
Enter the Dual-Piston Pump
 Figure 2: Two-piston pump designs: (a) Dual-piston pump with left piston in fill cycle and right piston in delivery cycle
and (b) accumulator-piston pump with top piston in delivery cycle and bottom piston in fill cycle. Arrows show direction of
motion of pistons; pressure and flow profiles are shown at right. See text for discussion.
|
The simple cure for the pulsating nature of the single-piston pump was to operate two such pumps so that when one piston was
filling, the other was delivering solvent. This is shown in the diagram in Figure 2a, where the left-hand piston is at the
end of its fill stroke and the right-hand piston is at the end of its delivery stroke. The solvent delivered from both pump
heads is combined into a single flow-stream. Because one piston is always in the delivery cycle, solvent flows continuously
to the column and pulses are minimized or eliminated. In practice, dual-piston pumps are configured so that the pistons are
mounted side-by-side in parallel. The dual-piston pump was developed early in the history of HPLC and remains as one of the
two most popular designs in use today.
Accumulator-Piston Pump
A later development in pump design is the accumulator-piston pump shown in Figure 2b. This also is called a tandem-piston
pump. In this design, the two pistons deliver solvent at two different flow rates. For example, if the pump of Figure 2b is
set to deliver 1 mL/min of solvent to the column, the top piston (shown at the end of the delivery stroke) pumps at 1 mL/min.
Meanwhile, the bottom piston fills at 2 mL/min. Next, the top piston fills at 1 mL/min. The 2 mL/min delivered by the bottom
piston is split so that 1 mL/min serves to fill the top piston and the other 1 mL/min flows directly to the column. In this
manner, solvent always flows to the column at 1 mL/min.