Often, one of the first signs of trouble with a liquid chromatography (LC) method is a failure to pass the system suitability
requirements for the method. In fact, early warning of potential problems is one of the primary purposes of a system suitability
test. As stated in the regulatory guidance,1 "The accuracy and precision of HPLC data collected begin with a well-behaved chromatographic system. The system suitability
specifications and tests are parameters that provide assistance in achieving this purpose." Unfortunately, sometimes we lose
sight of the purpose of a system suitability test and worry more about checking the boxes on a form than what the test is
trying to do. I am often asked what system suitability requirements are mandated by regulation. This is an interesting question,
because as far as I know, there are no firm requirements as to what parameters must be measured or what the minimum values
of these parameters must be. The closest thing that I can find to requirements is a document called "Reviewer Guidance: Validation
of Chromatographic Methods,"1 from the Food and Drug Administration's Center for Drug Evaluation and Research (FDA-CDER). This is a document intended
to give CDER reviewers of LC methods some guidelines about what to look for in a "good" method. Or as the document states,1 "The purpose of this technical review guide is to present the issues to consider when evaluating chromatographic test methods
from a regulatory perspective." Many workers feel that if their methods perform at least as well as the recommendations of
this document, they will be safe from regulatory criticism. This may or may not be a well-reasoned response. This month's
"LC Troubleshooting" discussion will centre on the recommendations of the CDER document, especially in terms of what it means
from a practical method performance standpoint.
The Recommendations
Table 1 summarizes four of the key parameters that can be used to evaluate method performance. These are standard measurements
that most workers make on a routine basis. Let's review the definitions of each of these, then look at some examples.
Retention factor, k. The retention factor (also sometimes called capacity factor, k') is a measure of the distribution of the analyte between the mobile phase and the stationary phase in the column in isocratic
(constant organic solvent) separations. It can be thought of as a way to measure retention in a manner that is independent
of column dimensions and flow-rate. Retention factor is calculated as where tR is the retention time of the analyte of interest and t0 is the column dead-time ("solvent front"). As has been discussed in other "LC Troubleshooting" columns (for example, see
reference 2), when developing a new method, we strive for 2 < k < 10, but usually will accept 1 < k < 20 for all the peaks of interest in a separation. Peaks with k < 1 can be compromised with real samples that contain large "garbage" peaks at t0. Methods with k > 2 also tend to be more robust to small changes in the chromatographic conditions.
Resolution, Rs. Resolution is the measurement of the separation of two peaks in a chromatogram and is a function of both the peak widths
w and retention times: