 John W. Dolan
|
Each week I get emails from various readers with questions or problems (see contact information at the end of this column).
I enjoy most of these and often they give me fodder for one of these "LC Troubleshooting" columns. This month I'd like to
look at one of those problems, because it can give us some insight into the effects that certain changes will make with a
liquid chromatography (LC) method.
The question went something like this: "I have a method that works well, but I am trying to scale it down so that I can save
acetonitrile. The method uses a 150 mm × 4.6 mm, 5-μm particle C18 column operated at 2 mL/min at 30 °C. This is an isocratic
method run with 40% acetonitrile and 60% of 0.1% formic acid. My sample is dissolved in mobile phase and I inject 20 μL. My
peaks come out at 1.5 and 1.65 min, with an overall cycle time of 4 min. The resolution requirement of Rs > 2.0 is obtained easily. I decided to switch to a smaller diameter column to reduce the acetonitrile consumption. So I switched
to a 150 mm × 2.1 mm, 5-μm C18 column that I found on the shelf and dropped the flow rate to 1 mL/min, because the column
is about half the diameter of the original one. Now I can barely pass the resolution requirements and sometimes fail. The
pressure is higher than before and the peaks come out earlier. This seemed so simple, but I must have done something wrong.
Can you help me?"
The Rule of One
This problem is a classic example of violation of the rule of one, which states, "change just one thing at a time." This is
the scientific method, and we should use it to help identify the cause-effect relationship of changes we make to the system.
As I see it, the column size, flow rate, and injection volume have changed and perhaps the column chemistry has, as well.
Let's look at some of the factors that we should consider when making a change such as the one mentioned previously. First, we need to make sure we have not made a chemistry change in the system. Based upon the question, I'm not sure if the
column chemistry is the same between the two columns. There was a time when everyone thought that all C18 columns were created
equal, but today, with literally hundreds of C18 columns to choose from, it might be more surprising if two are chemically
the same than if they are different. The 2.1-mm column should be from the same brand and line of packing material as the original
4.6-mm column. Because this wasn't mentioned specifically in the question, I want to make sure it is not overlooked. A second
way the chemistry of the column can change is if it has been used for other samples. A column that was "found on the shelf"
might or might not be new. If it is used, it still might be OK to use, but this decision should be based upon a column log
sheet that records column history and column testing. Any column with unknown history should, in my opinion, be filed in the
dumpster. Columns are consumable items with finite lifetimes, and it isn't worth the risk of creating problems with a method
by using a column with unknown history. Either of these changes, a different manufacturer's C18 material or a used column,
can mean a change in the column chemistry and, thus, a possible change in peak spacing — one of the possible reasons the resolution
requirements are hard to meet with the new column. For the moment, let's assume that the 2.1-mm column was from the manufacturer
and packing type and was new or like-new.